Dust Ingress Protection Testing for Product Enclosures and Components
The SC-010L Dust Testing Chamber is designed to evaluate the dust resistance and sealing performance of product enclosures and components under controlled test conditions. It is suitable for assessing protection levels IP5X and IP6X in accordance with applicable ingress protection standards.
By simulating dusty and sandy environmental conditions, the chamber helps manufacturers, research laboratories and quality assurance teams evaluate the effectiveness of enclosure seals and the durability of products exposed to airborne dust.
Applications include testing locks, automotive and motorcycle components, electrical instruments, seals and other products requiring protection against dust ingress.
Key Features
-
Suitable for IP5X and IP6X dust ingress protection testing.
-
1,000-litre internal chamber capacity.
-
7-inch programmable touchscreen controller with LCD display.
-
Adjustable dust blowing and vibration cycles.
-
Independent control of dust blowing, vibration and total test duration.
-
Programmable test sequences with automatic operation.
-
Stainless steel SUS304 internal chamber construction.
-
Integrated observation window with internal lighting.
-
Adjustable stainless steel sample rack.
-
Vacuum pressure indication.
-
Temperature-controlled dust preheating system.
-
USB data export and Ethernet communication capabilities.
-
Multiple safety protection functions, including overtemperature, overload and leakage protection.
Technical Specifications
| Parameter | Specification |
|---|---|
| Model | SC-010L |
| Internal chamber dimensions (D × W × H) | 1,000 × 1,000 × 1,000 mm |
| Nominal mesh aperture | 75 μm |
| Nominal metal mesh wire diameter | 50 μm |
| Dust particle size | 1–10 μm (more than 50% of particles ≤5 μm) |
| Dust concentration | 2–4 kg/m³ |
| Airflow velocity | ≤2 m/s |
| Dust blowing time | 1 second to 99 hours, adjustable |
| Vibration time | 1 second to 99 hours, adjustable |
| Test time setting | 0 seconds to 99 hours |
| Chamber temperature range | Room temperature +5°C to +60°C |
| Operating mode | Fixed-value operation |
| Standard test dust | Dry talcum powder, Portland cement or chimney ash |
| Net weight | Approx. 200 kg |
| Gross weight | Approx. 300 kg |
| External dimensions after packaging | Approx. 1,350 × 1,440 × 2,020 mm |
Chamber Construction
The chamber is constructed with a durable, powder-coated steel exterior and an imported SUS304 stainless steel internal liner.
Additional construction features include:
-
Single-door configuration with a silicone sealing structure for effective enclosure sealing.
-
Anti-condensation insulated observation window with integrated lighting.
-
Stainless steel SUS304 adjustable sample rack.
-
Heavy-duty castors for ease of positioning and movement.
-
Sealed door construction designed to provide reliable dust containment.
Dust Circulation and Heating System
The chamber incorporates a dedicated dust circulation system designed to distribute dust throughout the testing space.
Airflow is directed vertically from the top to the bottom of the chamber to promote an even distribution of test dust. A dual-motor arrangement supports dust circulation, while the integrated vibration system helps maintain dust suspension during testing.
A mica heating element provides controlled dust preheating, with timed heating functionality.
Programmable Control System
The SC-010L features a 7-inch touchscreen programmable controller with an LCD interface and configurable test parameters.
Operators can independently set and control:
-
Dust blowing cycles, including continuous or intermittent operation.
-
Vibration and rest periods.
-
Total test duration.
-
Temperature settings and operating sequences.
-
Automatic start and stop functions.
The controller also provides real-time monitoring, alarm notifications, fault diagnosis and power outage memory.
Data Recording and Communication
The control system supports historical data recording and curve playback, with the ability to export test records via USB for viewing or conversion to Excel format using the supplied software.
An Ethernet interface supports remote monitoring, program editing, historical data access and remote diagnostics. RS-485 and RS-232 communication interfaces are also available.
Safety Features
The chamber incorporates multiple protection functions to support safe operation, including:
-
Overtemperature protection.
-
Fan motor overheating protection.
-
Phase failure and reverse-phase protection.
-
Electrical leakage protection.
-
Overload and short-circuit protection.
-
Automatic shutdown in response to specified fault conditions.
Applicable Standards
The chamber references the following standards:
-
GB/T 4942-2021
-
GB/T 10485-2007
-
GB 4208-2017
-
GB/T 7000.1-2023
-
GB/T 2423.37-2006 – Environmental Testing for Electrical and Electronic Products: Sand and Dust Test
-
GB/T 2423.61-2018 – Environmental Testing: Large Specimen Sand and Dust Test
-
GB/T 38924.10-2020
-
GB 7001-2015
Please confirm the required test procedure and applicable standard before ordering, as chamber configuration and test requirements may vary.
Installation Requirements
The chamber should be installed on a level floor in a well-ventilated area, away from strong vibration, electromagnetic interference and flammable, explosive or corrosive materials.
Recommended clearance for installation and maintenance:
-
Side A: Minimum 600 mm
-
Side B: Minimum 1,200 mm
-
Side C: Minimum 600 mm
-
Side D: Minimum 900 mm
Recommended operating environment:
-
Ambient temperature: 5°C to 28°C
-
Relative humidity: ≤85% RH
-
Atmospheric pressure: 86–106 kPa
A forklift should be arranged to assist with unloading upon delivery.
Standard Supply
-
SC-010L Dust Testing Chamber
-
One-layer adjustable stainless steel sample rack
-
Certificate of Conformity
-
User operation manual
Important Operating Restrictions
The chamber must not be used for testing or storing flammable, explosive, corrosive or radioactive materials, highly toxic substances, or samples that generate highly toxic substances. Samples containing strong electromagnetic emission sources are also prohibited.
Enquire About the SC-010L Dust Testing Chamber
Contact Hylec Controls for further information, pricing and assistance with selecting a suitable dust testing solution for your application.